Critical

Oxygen supply failure / anaesthetic machine failure

Equipment emergency · Perioperative & Anaesthesia
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Loss of the oxygen supply or anaesthetic machine function is a life-threatening emergency that can rapidly cause hypoxia; the priority is to guarantee oxygenation and ventilation using an independent backup while maintaining anaesthesia.

Red flags

  • Low FiO2 alarm with falling SpO2
  • Inability to ventilate or deliver oxygen via the machine
  • Suspicion of crossed/contaminated pipeline gases
  • Simultaneous failure affecting multiple theatres (hospital supply failure)
  • Patient awareness risk if volatile delivery is lost without IV cover

Assessment

Differential diagnosis

  • Pipeline failure / crossed pipelines / pressure loss
  • Cylinder empty or not opened; regulator failure
  • Ventilator or bellows failure / power failure
  • Circuit disconnection, leak or soda-lime exhaustion/channelling
  • Vaporiser or flowmeter fault
  • Oxygen analyser/alarm malfunction masking a hypoxic mixture

Investigations

  • Inspired oxygen concentration (FiO2) analyser reading
  • Oxygen supply pressure gauges (pipeline and cylinder)
  • Capnography and airway pressure/volume monitoring
  • SpO2 trend
  • Circuit integrity check (disconnection, leaks)
  • Ventilator and power status indicators

Management

  • Recognise the problem (low FiO2/supply pressure alarm, failure to ventilate) and call for help
  • Turn ON the backup oxygen cylinder on the anaesthetic machine and check it is full
  • Disconnect the pipeline supply if a crossed pipeline or pipeline fault is suspected
  • Switch to a self-inflating bag (e.g. Ambu) with an independent oxygen cylinder to guarantee ventilation
  • Maintain anaesthesia intravenously (TIVA) since volatile delivery may be lost
  • Confirm oxygenation and ventilation with SpO2 and capnography
  • Use minimal/low fresh-gas flow or manual ventilation to conserve cylinder oxygen while help arrives
  • Call for additional cylinders and alert theatre coordinator, estates/engineering and other theatres
  • Consider aborting/pausing surgery at a safe point if supply cannot be secured
  • Identify and isolate the fault (machine, pipeline, hospital supply)
  • Keep the patient monitored and anaesthetised throughout the transition
  • Document and report the incident and quarantine faulty equipment

Drugs & doses

Propofol TCI or 100–200 micrograms/kg/min infusion to maintain anaesthesia (TIVA)
Remifentanil TCI / 0.05–0.25 micrograms/kg/min infusion for analgesia/anaesthesia
Midazolam 1–2 mg IV or ketamine 0.25–0.5 mg/kg IV as backup amnesia if infusions limited
Neuromuscular blocker as required to maintain relaxation
Oxygen from an independent full cylinder via self-inflating bag

Disposition

Complete or safely abandon surgery; continue in recovery/ICU as the patient's state dictates, report the incident, and remove faulty equipment from service until engineering clearance.

Complications & pitfalls

  • Hypoxia and hypoxic organ injury
  • Awareness under anaesthesia
  • Hypercapnia from inadequate ventilation
  • Delivery of a hypoxic gas mixture (crossed pipelines)
  • Haemodynamic instability

Exam pearls

  • A self-inflating bag with its own oxygen cylinder is the ultimate backup — have one in every theatre
  • If in doubt, disconnect from the machine/pipeline and ventilate independently
  • Maintain anaesthesia with TIVA when gas delivery is compromised to prevent awareness
  • Oxygen failure warning devices (and the oxygen ratio/pressure interlock) are key safety features
  • Perform and document the pre-use machine check to detect many faults before they cause harm

Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.