Cardiac resynchronisation & conduction system pacing
Bedside summary
Cardiac physiologic pacing aims to restore or preserve synchronous ventricular contraction. Traditional cardiac resynchronisation therapy (CRT) uses biventricular pacing — a right atrial lead, a right ventricular lead and a left ventricular lead in the coronary sinus — to correct dyssynchrony in heart failure with reduced ejection fraction and bundle branch block. Conduction system pacing (His-bundle pacing and left bundle branch area pacing) instead captures the native conduction system for a more physiological activation, and is increasingly used when a coronary sinus lead fails or to avoid right-ventricular pacing-induced dyssynchrony.
Red flags
- Rare ventricular tachycardia storm after CRT activation (proarrhythmia from altered repolarisation)
- Loss of biventricular capture in a device-dependent patient (lead dislodgement, high thresholds)
- Device-pocket infection or systemic infection seeding the leads
- Acute lead perforation / tamponade or coronary sinus dissection at implant
Assessment
Differential diagnosis
- Biventricular CRT (coronary sinus LV lead) — the established approach
- Left bundle branch area pacing (LBBAP) — preferred conduction system pacing option
- His-bundle pacing — physiological but technically demanding, higher/rising thresholds
- CRT-P (pacing only) vs CRT-D (added defibrillation)
Investigations
- ECG — QRS morphology and duration (left bundle branch block, pacing percentage); device interrogation for percent biventricular pacing
- Echocardiogram — LV ejection fraction and reverse remodelling as a measure of response
- Contrast venography before an upgrade (venous obstruction affects 10-30% of upgrade candidates)
- Assessment for non-response (~30%): adherence, progressive heart failure, suboptimal lead position or inadequate pacing capture
Management
- Select candidates with heart failure, reduced ejection fraction and electrical dyssynchrony (trial enrolment used LVEF <=35%, NYHA II-IV, LBBB with QRS >=140 ms in men / >=130 ms in women) on guideline-directed therapy
- Aim for near-100% biventricular/physiological pacing — optimise AV delay, and in AF with fast rates raise the lower rate or consider AV node ablation to guarantee capture
- Position the LV lead at the site of greatest electrical delay on the lateral/posterolateral wall, avoiding phrenic nerve capture and LV scar; quadripolar leads reduce phrenic stimulation
- For non-responders, re-optimise pacing vectors to the narrowest non-apical QRS, consider multipoint pacing, or convert to conduction system (left bundle area) pacing
Drugs & doses
Disposition
Complications & pitfalls
- Coronary sinus/coronary vein trauma, pneumothorax, pocket haematoma and phrenic nerve pacing
- LV lead dislodgement (~1.7%); failed LV lead placement in ~3-10%
- Device malfunction (~5%) and lead problems (~7%) over follow-up; infection ~1-2%
- Higher complication and mortality rates for upgrades than de novo implants
Exam pearls
- Greatest benefit is in left bundle branch block, wider QRS, non-ischaemic cardiomyopathy and women; little or no benefit (possible harm) with a narrow QRS
- Left bundle branch area pacing is now the preferred conduction system option — lower thresholds, higher implant success and better lead stability than His-bundle pacing
- A high pacing percentage matters — undertreated AF or frequent ectopy that steals capture undermines CRT
- CRT is also used to prevent pacing-induced cardiomyopathy in patients needing a high ventricular-pacing burden, including after AV node ablation
Resynchronisation only works if the patient actually gets it — secure LBBB-morphology candidates, place the LV lead well, and drive pacing towards 100%.
Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.