Pulmonary artery catheter & cardiac output monitoring
Bedside summary
The PA catheter measures RA/RV/PA pressures, wedge (PAWP), thermodilution cardiac output and mixed venous saturation — powerful physiology, but multicentre trials show no mortality benefit from routine use; modern practice favours less invasive tools (pulse-contour/peripheral thermodilution, echocardiography) with the PAC reserved for complex cases (mixed shock, pulmonary hypertension, post-cardiac surgery).
Red flags
- Haemoptysis after balloon inflation = possible PA rupture — call cardiothoracics immediately
- Persistent wedge trace = withdraw now
Management
- Insert via an 8Fr sheath (IJ/subclavian); balloon inflated only when advancing; watch waveform transitions RA→RV→PA→wedge; deflate after each wedge
- Continuous PA-trace vigilance: a permanently wedged trace risks pulmonary infarction — pull back; loops/knots avoided by noting insertion depth (~40 cm RIJ to PA)
- Interpret in sets: cardiac index, SVR, PAWP, ScvO₂/SvO₂ together — single numbers mislead
- Low output + high wedge = pump failure (inotrope/afterload work); low output + low wedge = hypovolaemia; high output + low SVR = distributive shock
- Prefer echo first in most shock: it answers “pump, tank or pipes?” fast and non-invasively
- Complications: arrhythmias (transient RV ectopy), complete heart block in pre-existing LBBB (have pacing available), PA rupture (balloon in distal PA — fatal risk), infection
Physiology in numbers — use selectively, interpret in patterns, and reach for the echo probe first.