Pulse oximetry & capnography — use and pitfalls

SpO₂ / EtCO₂ · Critical Care / ICU
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Oximetry: two-wavelength light absorption across a pulsatile bed estimates functional SaO₂. Capnography: continuous expired CO₂ — the waveform confirms tracheal intubation, tracks ventilation, perfusion and circuit integrity.

Red flags

  • Normal SpO₂ in a smoke-inhalation patient means nothing — measure COHb
  • Loss of the capnograph trace in a ventilated patient is an emergency, not an artefact

Management

  • Treat the patient, not the probe: check waveform quality before believing a bad number
  • Oximetry errors: carboxyhaemoglobin reads falsely HIGH (CO poisoning — use a co-oximeter gas), methaemoglobin drives readings toward ~85%, poor perfusion/vasoconstriction, nail varnish, movement, dyes
  • SpO₂ is blind to ventilation — a normal saturation on oxygen can hide severe hypercapnia; get a gas when in doubt
  • Capnography: flat trace after intubation = oesophageal tube or no circulation until proven otherwise
  • Falling EtCO₂: falling cardiac output/PE/hypovolaemia or circuit leak; rising: hypoventilation, malignant hyperthermia, bicarbonate
  • Use continuous waveform capnography for EVERY intubation, transfer and cardiac arrest (trend quality of CPR; sudden rise = ROSC)

Oximetry measures oxygenation only; capnography guards the airway, the ventilation and the circulation.