Oxygen delivery devices in children

Oxygen devices · Paediatric
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Match the device to the oxygen the child needs, and step up when they need more. Low-flow nasal cannula and simple masks cover modest needs; a non-rebreather gives the most oxygen short of positive pressure. Aim for SpO2 94% and above, not 100%.

Assessment

Investigations

  • Continuous pulse oximetry; give oxygen if SpO2 is 94% or less or there are two or more signs of distress
  • Reassess work of breathing and SpO2 after every step up
  • Keep the child in a position of comfort; humidify if oxygen is prolonged

Management

  • Low-flow nasal cannula 1-4 L/min (infants 2 L/min or less) gives about 24-40% FiO2 - for mild needs; heat and humidify above 2 L/min
  • Simple face mask 6-10 L/min (keep above 5 to avoid CO2 rebreathing) gives about 35-50%
  • Partial or non-rebreather mask with reservoir 10-15 L/min gives about 60-95% - the most oxygen short of positive pressure
  • Head-box/hood for infants gives about 80-90%; blow-by (unreliable, under 30%) only for the agitated infant who refuses a device
  • If the child needs more than 30% oxygen sustained, or is tiring, step up to high-flow nasal cannula or non-invasive / invasive ventilation

Drugs & doses

-

Disposition

Per severity; reassess frequently.

Blow-by oxygen is unreliable and a self-inflating bag cannot deliver it - use a flow-inflating bag or a proper device. A rising oxygen requirement is a signal to escalate support, not just to turn up the flow.

Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.