Anaesthetic machine check (AAGBI)

Machine safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

A structured pre-use check of the anaesthetic workstation and breathing system performed before every operating list to detect faults that could cause hypoxia, awareness or barotrauma.

Assessment

Investigations

  • Self-test per the machine manufacturer
  • Oxygen analyser calibration (21% and 100%)
  • Pipeline tug test and cylinder reserves
  • Two-bag test of the breathing system and ventilator
  • Capnography, vaporiser filling and scavenging checks

Management

  • Confirm the machine is plugged in with a reserve oxygen cylinder
  • Run the manufacturer self-test
  • Check pipelines by tug test and verify gauge pressures
  • Calibrate the oxygen analyser
  • Confirm vaporisers are seated, filled and leak-free
  • Perform the two-bag test on the circuit and ventilator
  • Check scavenging, suction, airway equipment, monitors and alarms

Drugs & doses

No drugs — equipment procedure
Verify an alternative means of ventilation (self-inflating bag) is present
Record the check before the first case

Disposition

Theatre; do not start without a completed check.

Perform a machine and breathing-system check before every list and after any change; always have a self-inflating bag available (AAGBI machine check 2012).