Capnograph sampling faults
Monitoring safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Capnography hardware problems such as a blocked or water-logged sampling line, loose connectors or an exhausted absorber can mimic or mask clinical events. Systematic checks restore a reliable trace.
Assessment
Investigations
- Inspect the sampling line for kinks and water
- Check connector seating (sidestream vs mainstream)
- Confirm the analyser is calibrated and zeroed
- Soda lime colour (raised baseline suggests exhaustion)
- Correlate with SpO2 and clinical signs
Management
- Confirm whether the problem is the patient or the equipment
- Check the sampling line for kinks, disconnection and water
- Clear or replace a blocked sidestream line and water trap
- Verify connectors and the airway adaptor are seated
- Recalibrate or zero the analyser if the baseline is wrong
- Replace exhausted soda lime if the baseline is raised
- Revert to clinical assessment while troubleshooting
Drugs & doses
No drugs — monitoring procedure
Spare sampling lines and water traps to hand
Replace absorber if inspired CO2 rises
Disposition
Theatre; never ignore a lost trace without excluding disconnection.
A raised baseline indicates absorber exhaustion or rebreathing; always exclude true disconnection before blaming the monitor (Miller).