Carbon dioxide rebreathing from absorber or valve failure
Rebreathing · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Exhausted soda lime or an incompetent unidirectional valve lets the patient rebreathe carbon dioxide, producing a rising baseline on capnography and hypercapnia. Recognising the trace prevents dangerous CO2 accumulation.
Assessment
Investigations
- Capnograph shows a baseline that does not return to zero
- Rising ETCO2 and inspired CO2
- Inspect CO2 absorbent colour change (exhaustion)
- Check unidirectional valves for sticking/incompetence
- Fresh gas flow setting
Management
- Increase fresh gas flow to flush the circuit
- Switch to an alternative circuit or self-inflating bag if needed
- Replace the exhausted CO2 absorbent canister
- Inspect and replace a faulty inspiratory/expiratory valve
- Confirm the capnograph baseline returns to zero
- Increase minute ventilation to clear accumulated CO2
Drugs & doses
100 percent oxygen; high fresh gas flow
Disposition
Continue once trace normal; tag machine for service.
An inspired CO2 above zero with a raised baseline is the signature of rebreathing — a stuck expiratory valve can cause it even with fresh absorbent.