Carbon monoxide and Compound A from CO2 absorbents

Absorbent toxins · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Desiccated carbon dioxide absorbent can generate carbon monoxide (with volatile agents) and, with sevoflurane, nephrotoxic Compound A. Fresh absorbent, adequate fresh gas flow and not leaving flows on overnight prevent exposure.

Assessment

Investigations

  • Review whether the machine was left with gas flowing (desiccated absorbent)
  • Raised carboxyhaemoglobin on co-oximetry if suspected
  • SpO2 (falsely normal with carboxyhaemoglobin)
  • Inspect absorbent for desiccation and heat
  • Monitor renal function after prolonged low-flow sevoflurane

Management

  • Use fresh, properly hydrated absorbent and change it regularly
  • Turn off all gas flow when the machine is not in use to prevent desiccation
  • Maintain adequate fresh gas flow during sevoflurane anaesthesia
  • If carbon monoxide exposure is suspected, give 100 percent oxygen and measure carboxyhaemoglobin
  • Avoid very low flows with potentially dry absorbent
  • Replace suspect absorbent canisters

Drugs & doses

100 percent oxygen if carbon monoxide exposure suspected

Disposition

Routine; equipment and renal follow-up if exposed.

A first case on a Monday after flows were left on over the weekend is the classic setting for carbon monoxide generation; keep absorbent hydrated and never leave fresh gas flowing on an idle machine.