Routine

Quantitative neuromuscular monitoring

TOF monitoring · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Objective (quantitative) monitoring of the train-of-four ratio to confirm full recovery before extubation, preventing residual paralysis and its airway and respiratory complications.

Assessment

Investigations

  • Acceleromyograph/electromyograph at the adductor pollicis
  • Train-of-four count and ratio
  • Post-tetanic count for deep block
  • SpO2 and clinical strength tests
  • Baseline calibration before relaxant

Management

  • Calibrate the monitor before giving the relaxant
  • Use train-of-four count to time intubation and reversal
  • Use post-tetanic count when no twitches are present
  • Guide reversal-agent choice and dose by the depth measured
  • Confirm a train-of-four ratio of at least 0.9 before extubation
  • Do not rely on clinical signs alone (they are insensitive)
  • Document the recovery ratio achieved

Drugs & doses

No drugs — monitoring technique
Targets: TOF ratio 0.9 or greater before extubation
Guides neostigmine vs sugammadex dosing

Disposition

Theatre / recovery; essential whenever relaxants are used.

A train-of-four ratio below 0.9 defines residual block and raises aspiration and hypoxaemia risk — confirm objectively (Miller).