Scavenging system safety

Equipment safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

The anaesthetic gas scavenging system removes waste anaesthetic gases to protect staff. A blocked active system can transmit negative or positive pressure to the patient, so correct function matters.

Assessment

Investigations

  • Confirm the scavenging is connected and switched on
  • Check tubing is not occluded or kinked
  • Reservoir bag/float indicator functioning
  • Air-break/relief valve present on active systems
  • Theatre waste-gas levels where monitored

Management

  • Connect the scavenging transfer tubing to the breathing system
  • Confirm the receiving system (active or passive) is working
  • Ensure the air-break prevents transmitted pressure
  • Check tubing for occlusion before use
  • Monitor the reservoir/float for correct flow
  • Avoid occluding an active system (barotrauma risk)
  • Report high theatre pollution for engineering review

Drugs & doses

No drugs — equipment procedure
Air-break interface protects the patient from applied pressure
Part of the pre-use machine check

Disposition

Theatre.

A blocked active scavenging line can apply harmful pressure to the circuit — the air-break interface is the key safeguard (AAGBI machine check).