Scavenging system safety
Equipment safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
The anaesthetic gas scavenging system removes waste anaesthetic gases to protect staff. A blocked active system can transmit negative or positive pressure to the patient, so correct function matters.
Assessment
Investigations
- Confirm the scavenging is connected and switched on
- Check tubing is not occluded or kinked
- Reservoir bag/float indicator functioning
- Air-break/relief valve present on active systems
- Theatre waste-gas levels where monitored
Management
- Connect the scavenging transfer tubing to the breathing system
- Confirm the receiving system (active or passive) is working
- Ensure the air-break prevents transmitted pressure
- Check tubing for occlusion before use
- Monitor the reservoir/float for correct flow
- Avoid occluding an active system (barotrauma risk)
- Report high theatre pollution for engineering review
Drugs & doses
No drugs — equipment procedure
Air-break interface protects the patient from applied pressure
Part of the pre-use machine check
Disposition
Theatre.
A blocked active scavenging line can apply harmful pressure to the circuit — the air-break interface is the key safeguard (AAGBI machine check).