Sudden loss of the capnograph trace
Flat ETCO2 · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
A sudden flat capnograph has three broad causes — the patient has no circulation, no ventilation, or the sampling equipment has failed. It must be treated as cardiac arrest until circulation is confirmed.
Assessment
Investigations
- Feel a central pulse immediately
- Check for disconnection, oesophageal tube or airway obstruction
- Inspect sampling line for water/kinking and the module for failure
- SpO2 and ECG
- Confirm chest movement and breath sounds
Management
- Feel a central (carotid/femoral) pulse at once
- If no pulse, start CPR and follow the cardiac arrest algorithm
- If pulse present, exclude ventilation failure: disconnection, obstruction, oesophageal intubation, severe bronchospasm
- Switch to hand ventilation with 100 percent oxygen
- If ventilation is adequate and pulse present, suspect sampling-line or monitor fault
- Replace sampling line or module and cross-check with a second monitor
Drugs & doses
100 percent oxygen
Adrenaline 1 mg IV per ALS algorithm if cardiac arrest confirmed
Disposition
Depends on cause — arrest call vs equipment swap.
Low rather than absent ETCO2 with ongoing CPR may indicate inadequate compressions or return of spontaneous circulation — use the waveform, not just the number.