Tracheal cuff pressure management

Airway safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Maintaining tracheal tube cuff pressure within a safe range protects against both aspiration (underinflation) and tracheal mucosal ischaemia (overinflation). A manometer gives objective control.

Assessment

Investigations

  • Cuff manometer reading
  • Audible leak at the set pressure
  • Capnography/ventilation for leak
  • Nitrous oxide use (diffuses into the cuff)
  • Signs of aspiration

Management

  • Inflate the cuff to just seal the airway
  • Measure the pressure with a manometer rather than by feel
  • Keep the pressure within the safe range
  • Recheck after nitrous oxide exposure (cuff expands)
  • Avoid overinflation to prevent mucosal ischaemia
  • Avoid underinflation to prevent aspiration and leak
  • Reassess after repositioning the patient

Drugs & doses

No drugs — airway procedure
Target cuff pressure 20-30 cmH2O (ideally about 25)
Release nitrous-oxide-induced overpressure with a manometer

Disposition

Theatre / ICU.

Cuff pressures above 30 cmH2O risk tracheal mucosal ischaemia; nitrous oxide diffuses in and raises the pressure (Miller).