Vaporiser faults
Equipment safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Vaporiser-related errors include wrong agent filling, overfilling, tipping, an empty vaporiser and interlock failure, any of which can cause overdose or awareness. Agent analysis is the key safeguard.
Assessment
Investigations
- Agent analyser (identity and concentration)
- Vaporiser fill level and seating
- Keyed filler and interlock check
- Inspired/expired volatile concentrations
- Depth-of-anaesthesia signs
Management
- Use agent-specific keyed fillers to prevent cross-filling
- Confirm the vaporiser is correctly seated with the interlock engaged
- Avoid overfilling and never tip a filled vaporiser
- Monitor inspired and expired agent concentrations continuously
- If tipping is suspected, flush with high fresh gas flows
- Replace a suspected faulty vaporiser
- Watch for awareness if the vaporiser runs empty
Drugs & doses
No drugs — equipment procedure
High fresh gas flow to flush a tipped vaporiser
Agent analyser confirms the correct volatile
Disposition
Theatre; remove and service a faulty unit.
A tipped vaporiser can deliver a dangerous overdose; continuous agent analysis detects wrong-agent and concentration errors (Morgan).